Beskrivelse
Weiss
Type: WK3-2200/40/5/V
Weiss Climate Test Chamber, Water-Cooled
Data for Temperature Testing:
Temperature range: -40°C to +180°C
Guodpfjw I Rxrsx Acvoh
Temperature change rate:
Heating 6.0 K/min
Cooling 11.5 K/min
Temperature deviation:
Temporal: ± 0.8 K
Spatial: ± 2.0 K
Heat compensation max. 8000 W
Data for Climate Testing:
Temperature range: +10 to +95°C
Dew point temperature range: +4 to +94°C with closed floor element
Humidity range: 10 to 95% RH
Temperature deviation, temporal: ± 0.3 K
Humidity deviation, temporal: ± 3% RH
Temperature uniformity:
Spatial: ± 1.0 K
Heat compensation max. 500 W
Test chamber volume: 2160 liters
Test chamber dimensions: Height 1115 mm, Width 1400 mm, Depth 1400 mm
External dimensions (lowered): Height 2318 mm, Width 1860 mm, Depth 3658 mm
External dimensions over shaker: Height 2710 mm
External dimensions max. lift height: 3038 mm
Interchangeable floor dimensions: Width 1600 mm, Depth 1504 mm
Operational Data:
Rated voltage: 3/N/PE AC 400V ±10% 50Hz
Rated power: 38.0 kW
Rated current: 63.0 A
Customer-side protection: 80 A slow-blow
Refrigerant: R 404A
Weight: approx. 2100 kg
For your safety as a buyer, please note the following information!
The following procedures are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. Refill with legally compliant refrigerant, if necessary
3. Leak test with certificate
4. After successful inspection, the chambers undergo a documented test run.
Condition: used
Scope of delivery: (see picture)
(Changes and errors in the technical data and details excepted!)
We will be pleased to answer further questions by telephone.
Annonsen ble oversatt automatisk. Oversettelsesfeil kan forekomme.
Type: WK3-2200/40/5/V
Weiss Climate Test Chamber, Water-Cooled
Data for Temperature Testing:
Temperature range: -40°C to +180°C
Guodpfjw I Rxrsx Acvoh
Temperature change rate:
Heating 6.0 K/min
Cooling 11.5 K/min
Temperature deviation:
Temporal: ± 0.8 K
Spatial: ± 2.0 K
Heat compensation max. 8000 W
Data for Climate Testing:
Temperature range: +10 to +95°C
Dew point temperature range: +4 to +94°C with closed floor element
Humidity range: 10 to 95% RH
Temperature deviation, temporal: ± 0.3 K
Humidity deviation, temporal: ± 3% RH
Temperature uniformity:
Spatial: ± 1.0 K
Heat compensation max. 500 W
Test chamber volume: 2160 liters
Test chamber dimensions: Height 1115 mm, Width 1400 mm, Depth 1400 mm
External dimensions (lowered): Height 2318 mm, Width 1860 mm, Depth 3658 mm
External dimensions over shaker: Height 2710 mm
External dimensions max. lift height: 3038 mm
Interchangeable floor dimensions: Width 1600 mm, Depth 1504 mm
Operational Data:
Rated voltage: 3/N/PE AC 400V ±10% 50Hz
Rated power: 38.0 kW
Rated current: 63.0 A
Customer-side protection: 80 A slow-blow
Refrigerant: R 404A
Weight: approx. 2100 kg
For your safety as a buyer, please note the following information!
The following procedures are carried out on our offered chambers in advance:
1. Functional check and replacement of necessary components
2. Refill with legally compliant refrigerant, if necessary
3. Leak test with certificate
4. After successful inspection, the chambers undergo a documented test run.
Condition: used
Scope of delivery: (see picture)
(Changes and errors in the technical data and details excepted!)
We will be pleased to answer further questions by telephone.
Annonsen ble oversatt automatisk. Oversettelsesfeil kan forekomme.
Tilbyder
Merk: Registrer deg gratis eller logg inn, for å få tilgang til all informasjon.
Send forespørsel
Telefon & Faks
+49 2861 ... annonser
Annonsen din har blitt slettet
Det har oppstått en feil